User Manual
Compare Glucose Monitor Test Result with Laboratory Results
The results obtained from the glucose monitor are plasma-calibrated. The result you obtain from your
glucose monitor may differ somewhat from your laboratory results due to normal variation. Monitor
results can be affected by factors and conditions that do not affect laboratory results in the same way.
To make an accurate comparison between monitor and laboratory results, follow the guidelines below.
Before the Lab Test
• If possible, fast at least eight hours before conducting a comparison test.
• Take your monitor to the lab.
While at the Lab
Make sure that samples for both tests are taken and tested within 15 minutes of each other.
• Wash your hands before obtaining a blood sample.
• Never use your glucose monitor with blood samples collected in a test tube.
• Use fresh capillary blood only.
CARE AND MAINTENANCE
• For regular maintenance, this device only needs to be wiped gently with a soft, dry cloth.
Never immerse this device or any components in water.
• Do not carry out repairs of any kind yourself. If a defect occurs, please contact your local
authorized distributor. Use only authorized parts and accessories.
• The cleaning and disinfection is absolutely necessary for the test procedure, because cleaning
can insure the monitor works well (for example, display will be clear to see after cleaning); and
disinfection can avoid the infection to you or to the other people, and the cross-infection.
• The monitor and lancing device should be cleaned and disinfected following each use.
• The monitor and the lancing device are for single‐patient use, if user test 6 times every day,
the monitor and lancing device should be cleaned and disinfected 6 times per day which equals
to 10950 number of cycles over the 5 year life of the devices. The monitor and lancing device
were validated for 11,000 cycles which could support up to 6 cleaning and disinfection cycles
per day.
• Below are the steps on how to clean and disinfect the monitor and lancing device.
1. After a test, clean and wash your hands.
2. Use wipes carefully to clean the entire external surface of the monitor.
3. Then wipe the entire external surface of the monitor with another wipe, keep the surface
wet for 2min.
4. Use the same method with the wipes to clean and disinfect the lancing device.
• Note:
- Wash hands thoroughly with soap and water after handling the monitor, lancing device, or test
strips.
- Only the surface of the monitor can be cleaned and disinfected with the disinfecting towelette.
Do not insert the disinfecting towelette into the test strip port and the metal connector, or else
the performance of the monitor may be affected.
- If you have any questions you can call your local customer service.
- If the monitor is being operated by a second person who is providing testing assistance to the
user, the monitor and lancing device should be cleaned and disinfected prior to use by the
second person.