TBP-6/0923 Blood Pressure Monitor Instruction Manual Please read all instructions carefully and retain for future use
Know your Device IndicaƟon for Use Warning BaƩery Warning 3 About Blood Pressure PreparaƟon 5 6 Setup your Device Seƫng of User, Year, Date and Time Before Measurement Measurement Using Memory 7 8 9 4 10 Care and Maintenance TroubleshooƟng 11 EMC DeclaraƟon 12 FCC Compliance 14 SpecificaƟons 15 Warranty InformaƟon 16
Know your Device Indications for Use This arm blood pressure monitor is intended to be used to measure blood pressure and pulse rate from the upper arm by using the oscillometric method, at home. This device is intended for use on adults only and is not for use on groups such as neonatal babies, children, pregnant women, or pre-eclamptic patients.
Warning (continued) • Do not immerse this device in water or clean with cleaning products, alcohol, or solvents. Carefully follow cleaning instructions provided. • Remove batteries if this device will not be used for three or more months. • Ensure that the device is used in the environment specified in the EMC declaration in this IM, otherwise, may result in improper operation. • Use of this device adjacent to or stacked with other device should be avoided because it could result in improper operation.
BLOOD PRESSURE INDICATOR CHART Systolic mmHg (Upper Number) (Based on 2017 ACC/AHA guidelines) 140 & above Stage 2 Hypertension * Red 130-139 Stage 1 Hypertension * Red 120-129 Elevated BP * Yellow Normal BP Green Below 120 Below 80 80-89 90 & above Diastolic mmHg (Bottom Number) *Consult A Physician When systolic and diastolic pressure fall into different categories, the higher pressure determines color code.
Cuff Display Set Button Start/Stop Button USB power supply port Memory Button Voice Battery Cover Air Socket Low battery Systolic pressure Blood pressure classification Diastolic pressure User number Memory Times Year/Date/Time IHB Average Value Unit Indicator Pulse rate Heartbeat
Set Up Your Device Figure 1 Ba ery Installa on 1. Take off the ba ery cover on the back of this device. (Figure 1) 2. Insert 3 AA alkaline ba eries according to the + and – markings. (Figure 2) 3. Snap the ba ery cover on, ensure it clicks into place. Figure 2 - + + - NOTE: When the low ba ery symbol appears on screen, turn off this device and follow the instruc ons above to replace ba eries. + CAUTION Never leave any low ba ery in the ba ery compartment as it may leak and cause damage to this device.
Setting of User With device turned off, press and release “SET” bu on to show the current user. Press “SET” bu on to toggle between User 1 and User 2. To select user, press “START/STOP” to confirm and then the device turns off. Setting of Year, Date, Time (12-hour/24-hour System) 1. With device turned off, press and hold “SET” bu on un l flashing year digits occur. 2. Press “MEM” to increase one year, or press and hold “MEM” to increase years. Press “SET” to confirm and then flashing month digits occur. 3.
Before Taking a Measurement For most accurate results, measurements should be taken in the same circumstances and at the same me every day. This device can be applied to either arm, but preferably the le . Use the same arm for each measurement. Any blood pressure recording can be affected by the posi on and physiological condi on of the subject, as well as other factors such as background noise, stress, talking and body movement etc. Avoid taking measurements while under stress.
Measurement (con nued) 4. While si ng upright in a chair with your feet flat on the floor, press “START/STOP". Symbols on screen will flash and cuff will begin to inflate. Do not move, ac vate arm muscles, or talk during measurement. 5. Measurement can be stopped at any me by pressing “START/STOP". 6. When a pulse is detected, a heart symbol will flash on screen. 7. When measurement is complete, the cuff deflates automa cally and your systolic, diastolic, and pulse rate will display on screen. 8.
Care and Maintenance Care and Maintenance For regular maintenance, this device only needs to be wiped gently with a so , dry cloth. Never immerse this device or any components in water. Do not carry out repairs of any kind yourself. If a defect occurs, please contact your local authorized distributor. Use only authorized parts and accessories.
Guidance and Manufacturer’s DeclaraƟon - ElectromagneƟc Immunity this device is intended for use in the electromagneƟc environment specified below. The customer or the user of this device should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level ± 8 kV Electrostatic contact discharge (ESD) IEC 61000-4-2 ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air Electromagnetic Compliance Environment - Guidance Level ± 8 kV Floors should be wood, contact concrete, or ceramic tile.
This device complies with Part 15 of the FCC rules. Opera on is subject to the following two condi ons: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired opera on. Cau on: Changes or modifica ons to this device not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
1 Model TBP-6/0923 2 Measurement Method Oscillometric 3 Display LCD 4 Measurement Range Pressure: 0 to 299 mmHg 5 Accuracy Pressure: Within ± 3 mmHg 6 Inflation Automatic inflation by pump 7 Deflation Automatic rapid deflation 8 Pressure Detection Semiconductor pressure sensor 9 Memory 2 persons X 90 measurements each 10 Power Supply 3 AA alkaline batteries (not included) Battery Life About 6 months when used once a day Pulse: 40 to 180 beats/minute Pulse Rate: Within ± 5% of
WARRANTY DURATION: All materials and workmanship are warranted to the original consumer purchaser for a period of ninety (90) days from the original purchase date. WARRANTY COVERAGE: This product is warranted against defec ve materials or workmanship. This warranty is void if the product has been damaged by accident, in shipment, unreasonable use, misuse, neglect, improper service, commercial use, repairs by unauthorized personnel or other causes not arising out of defects in materials or workmanship.