TTM-6/0740 Low-Frequency Therapy Instrument Instruction Manual Please read all instructions carefully and retain for future use
Know your Device Indication for Use Medical Disclaimer Precautions 3 Important Safety Information Battery Warning 4-5 Preparation Package Content Parts About Therapy Machine 6 Set up your Device Getting Start After Used 7 Instruction of Use 8 Care and Maintenance Troubleshooting 9 EMC Declaration 10-12 Specification 13 Warranty Information 14
Know your Device Indications for Use The Low-Frequency Therapy Instrument is a product that adopts modern electronic science and technology to produce low-frequency electrical pulse and transmit that through the skin to the underlying peripheral nerves by electrodes, in order to reach the therapeutic aim. This device used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, leg and foot, due to strain from exercise or normal household and work activities.
Important Safety Information 1. Never use this product along with other medical electronic instruments, such as cardiac pacemaker, artificial heart-lung which are used to maintain life, and electrocardiograph. Otherwise it may cause danger. 2.
13. 14. 15. 16. transthoracically and transcerebrally; over swollen, infected, or inflamed areas or skin eruptions; over, or in proximity to, cancerous lesions. Do not arbitrarily disassemble, repair and reform this instrument, otherwise it might cause malfunction or electric shock accidents. 17. Stimulation should not be applied over the neck or mouth.
Package Content Low-Frequency Therapy Instrument Machine Pair of Electrode Pads Connecting Cable Storage Pouch 2 AAA Batteries (Included) Parts About Therapy Machine Therapy Machine User can select appropriate treatment mode and other parameters, such as intensity, through function buttons (shown below) on the host and know about the current status through LCD at any time. Treatment Mode 5 Full-automatic treatment mode and 5 manual treatment mode which satisfy to different demands.
Set Up Your Device 1. Open battery cover and insert 2 AAA batteries into the battery compartment according to polarity. Caution: When the intensity stimulation is very weak, it means the battery is low and a new battery should be replaced. 2. Insert one end of two-core connecting cable to machine socket and connect the other to electrode pad. Caution: The connecting line can only be used to connect electrode patch and the host; it is prohibited to be used for other purposes.
Set Up Your Device Instruction for Use 1. Press “On/+” button to power on and start working procedure, the instrument enters into automatic combination mode 5 (intensity is 1st level, timing for 15 minutes), LCD displays corresponding contents. 2. Press “On/+” button after the instrument has been powered on to enhance intensity which is divided into 9 levels. Intensity is initially set to 1st level (the weakest) by default, LCD will display corresponding contents. 3.
For regular maintenance, this device only needs to be wiped gently with a soft, dry cloth. Never immerse this device or any components in water. • Do not carry out repairs of any kind yourself. If a defect occurs, please contact your local authorized distributor. Use only authorized parts and accessories. • Keep the surface of electrode patches clean, or else the viscidity of electrode pad will be lower.
EMC Statement This device needs special precautions regarding EMC and need to be installed and put into service according to the EMC information provided in the accompanying document. Portable and mobile RF communications equipment can affect this device.
Immunity Test Electrostatic discharge (ESD) IEC 61000-4-2 IEC 60601 Test Level + 8 kV contact + 2 kV, + 4 kV, + 8 kV, + 15 kV air Compliance Level + 8 kV contact + 2 kV, + 4 kV, + 8 kV, + 15 kV air Electric fast transient/brust IEC 61000-4-4 + 2 kV for power supply lines + 1 kV for input/output lines Not applicable Surge + 1 kV line to line IEC 61000-4-5 + 2 kV line to earth Electromagnetic Environment - Guidance Floors should be wood, concrete, or ceramic tile.
Immunity Test Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-6 IEC 60601 Test Level 3 V/m 150kHz to 80MHz 6 Vrms in ISM bands 10 V/m 80MHz to 2.7GHz Compliance Level Not applicable Electromagnetic Environment - Guidance Portable and mobile RF communications equipment should be used no closer to any part of the models including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. 10 V/m 80MHz to 2.
Model TTM-6/0740 Dimensions – Therapy Machine 147mm (L) x 59mm (W) x 27.5mm (D) Dimensions – Electrode Pads 83mm x 59mm x 6mm Dimensions – Connecting Cable Cable Length: 1.2m Weight – Therapy Machine 2.7oz Power 2 AAA (Included) Number of Channels / Modes 1 / 10 Safety category Type BF Output Specification Output Intensity Level 9 Levels Waveform and Shape Pulsed symmetric, Biphasic, Square wave Maximum Output Voltage (±10%) 36.5V @ 500 Ω / 54.5V @ 2k Ω / 56.
WARRANTY DURATION: All materials and workmanship are warranted to the original consumer purchaser for a period of ninety (90) days from the original purchase date. WARRANTY COVERAGE: This product is warranted against defective materials or workmanship. This warranty is void if the product has been damaged by accident, in shipment, unreasonable use, misuse, neglect, improper service, commercial use, repairs by unauthorized personnel or other causes not arising out of defects in materials or workmanship.