Manual

16
SPECIFICATIONS
PRODUCT SPECIFICATIONS
Materials of Construction This product contains no latex.
Top Frame Extruded aluminum with powdercoated or anodized surfaces.
Table Top .5” (1.3 cm.) Polycarbonate
Patient Comfort 1” (2.5 cm.) upholstered foam pad
Table Length 78” (198 cm.)
Table Width 24” (61 cm.)
Height Range 24”-40” (61-102 cm.) standard
Table Top Travel Range ±11.5” (29 cm)
Trendelenburg Tilt ±14°
Weight 390 lbs. (177 kg.)
Shipping Weight 526 lbs. (239 kg.)
Lifting Capacity 450 lbs. (205 kg.)
Aluminum Filtration
Equivalent of Table top
TABLE TOP
100 kvp, Hvl of 3.6 mm = 1.62 (1.74 max)
1” (2.5 cm.) TABLE TOP PAD
100 kvp, Hvl of 3.6 mm = .88 (1.07 max)
2” (5 cm.) TABLE TOP PAD
100 kvp, Hvl of 3.6 mm = 1.47 (1.76 max)
This product complies with the United States Department of Health and Human Services radiation performance
standards, 21 CFR Subchapter I, in effect at the time of manufacture for radiographic tables.
ENVIRONMENTAL CONDITIONS
Conditions Temperature Humidity Atmospheric Pressure
Normal Use 50° (10°C) to 104° (40°C) 20% to 60% RH 98 to 105 kPa
Storage & Transport -20° (-29°C) to 135° (57°C) 20% to 95% RH 98 to 105 kPa
ELECTRICAL SPECIFICATIONS
Designed for: North America Europe Japan
Input Service 120 VAC/15 amp/60 Hz 220 VAC/10 amp/50/60 Hz 100 VAC/20 amp/50 Hz
Current Draw 5.8 amps 2.9 amps 4.8 amps
Maximum Momentary Cur-
rent Consumption
6.0 amps 3.0 amps 7.2 amps
Voltage Output to Actuators 24 VDC 24 VDC 24 VDC
Electric Shock Protection Class 1 Equipment Class 1 Equipment Class 1 Equipment
Tabletop Applied Part
Type B Applied Part
Type B Applied Part
Type B Applied Part
Ingress Protection Rating IPX0 IPX0 IPX0
Made of Operation Intermittent Operation
MAX 2 minutes ON
18 minutes off
Intermittent Operation
MAX 2 minutes ON
18 minutes off
Intermittent Operation
MAX 2 minutes ON
18 minutes off