Owner manual

FL-400 Fiber Probe Installation and Operation Instructions
F400-20CM-HT-01-201106 3
Care of the Probe
Follow these guidelines to ensure that the probe remains functional for as long as possible:
Handle the probe with care.
Do not drop the probe. Dropping the probe may cause permanent damage.
Clean the probe with distilled water or light detergents. Mild sonication is recommended.
Do not place probe tip in a cool cleaning solution while tip is hot. This could fracture the
silica fiber in the probe. Allow the probe to cool for a minimum of 10 seconds before placing
the probe in a cleaning solution.
Probe Specifications
Specification Value
Fiber core diameter
400
m
Fiber core/cladding Fused silica core and doped, fused silica cladding
Fiber jacketing Gold
Fiber type 1 single-strand, multimode fiber
Wavelengths covered 300 to 800 nm
Probe sleeve (ferrule) Nickel
Probe dimensions 17.78 cm length, 20 gauge probes with 0.902 mm OD
Temperature range (tip of probe)
–269
C to 700 C
Numerical aperture 0.22
Fiber termination SMA 905