User Manual

Operators Guide
18
3. Carefully pull the battery out of its compartment as far as the wiring will allow.
4. Pull the connector off the terminals of the old battery and push it onto the terminals of
the new battery. For longer battery life it is recommended that you use a 9 V lithium
battery. However, a standard 9 V battery can also be used.
5. Replace the battery in its compartment, making sure that the wiring is not twisted or
pinched.
6. Replace the battery cover and tighten the three slotted screws.
7. Press the display actuator button and verify that the digital display and bubble level are
lit for approximately 15 seconds.
Balance mechanism calibration
The digital display indicates the setting of the balance mechanism on a scale of 0 (minimum
setting) to 182 (maximum setting). In the unlikely event of this system requiring calibration,
proceed as follows:
1. Level the platform and engage the tilt lock pin.
2. Press and hold the Display Actuator button for 20 seconds to enable calibration mode.
Figure 5 Battery replacement
Battery compartment
cover
9V PP3 battery
slotted screws