Manual

12
hhmmsshhmmss
hhmmsshhmmss
hhmmss
TIME TIME
TIME TIME
TIME Set current time to "hh: mm: ss", hh is between 0 - 23.
TIMETIME
TIMETIME
TIME Print current time.
mmddyy DATEmmddyy DATE
mmddyy DATEmmddyy DATE
mmddyy DATE Set current date 'mmddyy'.
DATEDATE
DATEDATE
DATE Print current date.
%%
%%
% Print current % ref.
##
##
# Print current ref., any function.
ACAC
ACAC
AC Abort calibration /Test.
CC
CC
C Start an auto calibration.
EE
EE
E Go to initial state of current function.
FF
FF
F Print current function.
MM
MM
M Same as mode button.
xDxD
xDxD
xD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).
IDID
IDID
ID Print current ID string.
xxxxIDxxxxID
xxxxIDxxxxID
xxxxID Enter ID string (xxxx limit to 8 characters).
RS232 COMMAND TABLERS232 COMMAND TABLE
RS232 COMMAND TABLERS232 COMMAND TABLE
RS232 COMMAND TABLE
Field: Mode Stab CR LF
Length: 5 1 1 1
blank if stable
“ ? ” if unstable
CommandCommand
CommandCommand
Command
CharacterCharacter
CharacterCharacter
Character
DescriptionDescription
DescriptionDescription
Description
? ?
? ?
?
Print current unit.
0 = minimum level
1=
2 = maximum level
xIxI
xIxI
xI Set averaging level to "x", where x = 0, 1,
2
GRAMS, DWT, CARAT, OZ AV, OZ T, GRAIN,
TAEL 1, TAEL 2, TAEL 3, MOMME,CUSTOM UNIT
xMxM
xMxM
xM Places balance in mode "x",
where x = 1 to 11 (see table).
If unit is not already enabled,
command will be ignored.
1 = grams
2 = pennyweight
3 = carats
4 = avoidupois ounces
5 = troy ounces
6 = grains
7 = taels
8 = mommes
11 = custom unit
PP
PP
P Print display data
When “numeric only” display
data is selected for output in
the RS232 menu, the Mode
field is not output.
Field: Weight Mode Stab CR LF
Length: 10 1 5 1 1 1
Same as ?
command
Displayed weight sent right justified
w/lead zero blanking.
Nine characters include:
decimal point (1)
weight (7 max))
polarity (1): blank if positive
“ - ” if negative
INSTALLATION