Calibration Menu ProtectionCalibration Menu Protection
Calibration Menu ProtectionCalibration Menu Protection
Calibration Menu Protection
NOTESNOTES
NOTESNOTES
NOTES: 1. Calibration may be locked out to prevent unauthorized personnel
from changing calibration. If calibration has been locked out, you
can only access Test.
2. To lock out calibration menu, after calibration, refer to the section
titled Menu Lock-Out Protection.
Auto CalibrationAuto Calibration
Auto CalibrationAuto Calibration
Auto Calibration
Auto calibration is used when it is desired to calibrate the balance automatically.
Proceed as follows:
1. Make sure there is no load on the pan and close the chamber doors.
2.
.
3. .
4.
.
CALIBRATION MENUCALIBRATION MENU
CALIBRATION MENUCALIBRATION MENU
CALIBRATION MENU
Analytical Plus balances features
Auto Auto
Auto Auto
Auto
,
USERUSER
USERUSER
USER
and
TEST TEST
TEST TEST
TEST
calibration methods.
AutoAuto
AutoAuto
Auto
is a method where the balance calibrates itself using internal calibrated masses.
USER USER
USER USER
USER
is a method where the balance
can be calibrated using an external mass of
known value by entering that value into the balance.
Test Test
Test Test
Test
allows the stored
calibration data to be tested against the internal mass being used for the test. The
following figure illustrates the sequence in which submenus appear on the Calibration
menu. Item shown bolded is a default setting.
CALIBRATION MENUCALIBRATION MENU
CALIBRATION MENUCALIBRATION MENU
CALIBRATION MENU
AUTO
USR
Enter #
TEST
END, Calibration
NOTE: DO NOT DISTURB THENOTE: DO NOT DISTURB THE
NOTE: DO NOT DISTURB THENOTE: DO NOT DISTURB THE
NOTE: DO NOT DISTURB THE
BALANCE WHEN -C- BALANCE WHEN -C-
BALANCE WHEN -C- BALANCE WHEN -C-
BALANCE WHEN -C-
IS DISPLAYED. IS DISPLAYED.
IS DISPLAYED. IS DISPLAYED.
IS DISPLAYED.
MENUS