Manual

67
Error Codes ListError Codes List
Error Codes ListError Codes List
Error Codes List
The following list describes the various error codes and which can appear on the
display and the suggested remedy.
Data ErrorsData Errors
Data ErrorsData Errors
Data Errors
0.0 Internal data errors. If error persists, the balance must be serviced.
1.0 Internal data errors. If error persists, the balance must be serviced.
Tare ErrorsTare Errors
Tare ErrorsTare Errors
Tare Errors
2.0 Illegal tare operation or balance is unable to stabilize within time limit after
taring. Environment is too hostile or balance needs recalibration.
Calibration ErrorsCalibration Errors
Calibration ErrorsCalibration Errors
Calibration Errors
3.0 Incorrect or no calibration weight used for User calibration. Recalibrate with
correct weights.
5.0 Auto calibration failed. Environment is to hostile, recalibrate the balance.
RS232 ErrorsRS232 Errors
RS232 ErrorsRS232 Errors
RS232 Errors
4.4 RS232 buffer is full (if installed). May occur if no printer or computer is
connected to the interface. To clear buffer, turn balance off or enter Print menu
and select END.
4.5 Function is disabled by the Lock switch.
User ErrorsUser Errors
User ErrorsUser Errors
User Errors
7.0 User entry out of range (custom factor, non-pH, non-pL, etc...)
7.1 Bad percent (%) mode, sample too low.
7.2 Number outside of display capacity.
Over-Under Load ErrorsOver-Under Load Errors
Over-Under Load ErrorsOver-Under Load Errors
Over-Under Load Errors
8.0 Hardware error causing an internal weight signal which is too low. Check if
platform or platform support is off. If not, the balance must be serviced.
8.2 Internal power on error. Turn the power off, then turn it back on.
8.3 Hardware error caused by an internal weight signal which is too high. Check
load on the platform which may be excessive. If error persists, the balance must
be serviced.