99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.
Valor™ 2000W Series EN-1 1. INTRODUCTION This manual contains installation, operation and maintenance instructions for the Valor™ 2000W Series. Please read the manual completely before using the scale. 1.1 Safety Precautions Please follow these safety precautions: Verify that the AC Adapter input voltage matches the local AC power supply. Do not drop loads on the platform. Do not place the scale upside down on the platform. Disconnect the scale from power supply when cleaning.
EN-2 Valor™ 2000W Series 2.3 Selecting the Location Use the scale on a clean, firm and flat surface. Avoid locations with excessive air current, vibrations, heat sources, or rapid temperature changes. 2.4 Leveling the Equipment Adjust the leveling feet so the bubble is centered in the circle. Be sure the equipment is level each time its location is changed. Level Bubble Adjustable Feet Figure 2-2. Level Indicator 2.5 Power The AC Adapter is used to power the scale when battery power is not needed.
Valor™ 2000W Series EN-3 2.5.1 Battery Power: The scale can be used on AC power immediately. Allow the battery to charge for 12 hours before using the scale on battery power. The scale will automatically switch to battery operation if there is a power failure or the power cord is removed. With AC power, the scale is constantly charging, so the battery charge indicator (see item 8 in table 3-2) will remain lit. The scale can be operated during charging, and the battery is protected against overcharging.
EN-4 Valor™ 2000W Series 3. OPERATION – V22PW 3.1 Controls Figure 3-1.
Valor™ 2000W Series EN-5 TABLE 3-2 Display Symbols Item Description Item Description 1 NET symbol 7 Dynamic symbol* 2 Center of Zero symbol 8 Battery charge symbol 3 Negative symbol 9 Pound, Ounce, Pound:Ounce symbol 4 Stable weight symbol 10 Gram, kilogram symbol 5 Preset Tare symbol* 11 Pieces symbol* 6 Pointer symbols* 12 Percent symbol* Note: * Not Used Figure 3-3. Valor 2000W – V22PW Rear display. 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button.
EN-6 Valor™ 2000W Series 3.4 Initial Calibration When the scale is first installed it should be calibrated to ensure accurate results. Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 3-3. Press and hold Cal for 5 seconds, the display will show [Cal]. When the button is released, the display will show [SpaN]. Press On/Zero to begin the span calibration. [--C--] blinks while zero reading is stored.
Valor™ 2000W Series EN-7 4. OPERATION – V22XW 4.1 Controls Figure 4-1.
EN-8 Valor™ 2000W Series LED Indicators Figure 4-1. Valor 2000W – V22XW Rear display. The colored LED indicators on the right side of the control panels are used in Checkweigh mode (section 4.5) and will light up according to the following rules: (Red) Loads > Upper limit (Green) Loads ≥ Lower limit and ≤ Upper limit (Yellow) Loads < Lower limit Display See section 3.1 for information about the display and it’s symbols. 4.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button.
Valor™ 2000W Series EN-9 4.4 Weigh Mode 1. 2. 3. Press and hold Mode until [wJeIGH] (Weigh) is displayed. If required, place an empty container on the pan and press Tare. Add sample to the pan or container. The display shows the weight of the sample. 4.5 Checkweigh Mode This mode sets low and high weight limits for portion control processes. 1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear references) will then display.
EN-10 Valor™ 2000W Series 4.6 Menu Settings The User Menu allows the customizing of scale settings. Note: Valor 2000W – V22PW does not have these Menus. 4.6.1 Menu Navigation User Menu: Menu: C.A.L Span Lin Menu GEO Items: End M.O.d.E Reset Check End U.n.i.t kg g oz lb lb:oz End E.n.d To Enter the Menu Mode Press and hold Menu until [mMeNU] (Menu) is displayed. When released the first submenu [C.a.l] (Cal) will be shown. Press Yes to enter the displayed sub-menu or press No to advance to the next.
Valor™ 2000W Series EN-11 4.6.4 Unit Menu Enter this menu to activate units so they will be accessible with the Units button. The units in the menu must be turned “on” to be active. Note: Available units vary by model and local regulations. 4.6.5 End Menu Press ‘Yes’ to advance to the Calibration menu. Press ‘No’ to exit the menu and return to the current application mode. TABLE 4-2.
EN-12 Valor™ 2000W Series 5. MAINTENANCE 5.1 Cleaning The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panels. 5.2 Cleaning the Plastic Pan 5.2.1 Uninstalling and cleaning the plastic pan Please follow the steps below to clean the plastic pan: 1. Remove the stainless steel pan 2. Remove the four thumb screws holding the battery cover. 3.
Valor™ 2000W Series EN-13 5.3 Troubleshooting The following table lists common problems and possible causes and remedies. If the problem persists, contact OHAUS or your authorized dealer.
EN-14 Valor™ 2000W Series 6.1 Specifications TABLE 6-1 V22PWE1501T V22XWE1501T 1.5 kg x 0.0002 kg 1500 g x 0.2 g 3 lb x 0.0005 lb 48 oz x 0.01 oz 7500 0.2 g ± 0.2 g V22PWE3T V22PWE6T V22PWE15T V22XWE3T V22XWE6T V22XWE15T 3 kg x 0.0005 kg 6 kg x 0.001 kg 15 kg x 0.002 kg 3000 g x 0.5 g 6000 g x 1 g 15000 g x 2 g Capacity x Readability (Max x d) 6 lb x 0.001 lb 15 lb x 0.002 lb 30 lb x 0.005 lb 96 oz x 0.02 oz 240 oz x 0.05 oz 480 oz x 0.
Valor™ 2000W Series EN-15 6.3 Compliance Compliance to the following standards is indicated by the corresponding mark on the product. Mark Standard This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive 2006/95/EC. The Declaration of Conformity is available online at europe.ohaus.com/europe/en/home/support/compliance.aspx. AS/NZS CISPR 11 CAN/CSA-C22.2 No. 61010-1-12 UL Std. No.
EN-16 Valor™ 2000W Series FCC Note This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.