User Manual

EN-6 Valor 2000W Series
3.4 Initial Calibration
When the scale is first installed it should be calibrated to ensure accurate results.
Before performing the calibration, be sure to have the appropriate calibration
weights as listed in table 3-3.
Press and hold Cal for 5 seconds, the display will show [Cal]. When the button is
released, the display will show [SpaN]. Press On/Zero to begin the span calibration.
[--C--] blinks while zero reading is stored. Next, the display shows the calibration
weight value. Place the specified calibration mass on the pan and press On/Zero.
[--C--] blinks while the reading is stored. The scale returns to the previous
application mode and is ready for use.
The message [Err 3.0 Cal] will be displayed if these calibration steps are not
followed or if the wrong weight was used.
The calibration process can be aborted by turning the scale off.
TABLE 3-3
Required Span Calibration Mass (sold separately)
Max
Mass
Max
Mass
1500 g
1 kg / 3 lb
6000 g
5 kg / 15 lb
3000 g
2 kg / 6 lb
15000 g
10 kg / 30 lb
Note: If current unit is lb or oz, the scale will be calibrated by pound weight.
3.5 Unit Menu
Follow below instructions to enter the service menu to set the displayed unit.
With the scale off and LFT off, extend press both the On/Zero and Tare keys at the
same time for more than 8 seconds until [rAmMP] is displayed. Press the Tare key
through the succeeding displays: [rAmMP]*, [LIN]*, [SPAN]*, [CAP]*, [Dius]* until [s.UNIt]
is displayed.
When
[s.UNIt]
is displayed, press the On/Zero key.
[UNIt
kg
]
will be displayed. Press the Tare key until the active (power on)
unit is displayed.
When the active unit is displayed, press the On/Zero key.
[ON
XX
]
is displayed. Press Tare to switch to
[Off
XX
]
, then press the
On/Zero key**.
The next unit
[UNIt
xx
]
is displayed. Press the Tare key until the desired
unit is displayed.
When the desired unit is displayed, press the On/Zero key.
[Off
XX
]
is displayed. Press the Tare key to switch to
[ON
XX
]
, then press the
On/Zero key.
Press the Tare key until
[End]
is displayed, then press the On/Zero key.
Press the Tare key through the succeeding displays:
[e.paND]
*,
[reset]
*
until
[End]
is displayed.
When
[End]
is displayed, press the On/Zero key to exit to weighing mode.
Notes:
*Do not change any of these settings (by pressing the On/Zero key). This may affect
the correct set up of the scale.
**The active unit must be turned off first before another unit can be turned on.