I. FLAT button: Click the button, first the LIFT actuator goes to the low limited switch,
then the Head and Foot actuator will start up simultaneously and keep running until
arriving to the FLAT position. Running can be stopped by clicking any button.
J. MEMORY A button: Click the button, first the LIFT actuator goes to the Memory A
position, then the Head and Foot actuator will start up simultaneously and keep running
until arriving to the Memory A position. Running can be stopped by clicking any button.
K. MEMORY B button: Click the button, first the LIFT actuator goes to the Memory
B position, then the Head and Foot actuator will start up simultaneously and keep running
until arriving to the Memory B position. Running can be stopped by clicking any button.
The blue LEDs for backlight are on as long as any button is pressed. IF all buttons are
released the LEDs will automatically turn off three seconds later.
The blue LED for power is always on as long as the power on.
The below positions can be reset by the end user.
The Zero gravity position (only both Head and Foot can be reset, but not for the
Lift);
The Memory A position;
The Memory B position;
The procedure to set these special positions is below:
Step 1: Click the button to flat the bed;
Step 2: Adjust the Lift and Head and Foot position to a desired position by
pressing the corresponding button;
Step 3: Press and hold the button for 3sec, backlight LED start flashing;
Step 4: Release the button then click one of or or
button before LED stop flashing. Then the Flashing stops immediately to indicate
the corresponding operating has completed successfully.
ALL of the Zero gravity position and the Memory A position and the Memory B