The control box is designed for using the SMPS of 29VDC/ around 2A as the power. It is
necessary to use the SMPS provided by us when normal operation.
In the case of the AC power utility failure, the batteries for backup of 18VDC can be used as the
power. It is only for flatting the bed plate without load if the batteries used as the power and
each one of the batteries must be new with 9V voltage. You can press and hold the emergency
button on the body of the control box to flat the bed. The movement sequence is that the Foot
actuator goes down until arrives to the limit the Head one goes down. If release the button, all
movement will stop immediately including the massage motors.
It isn’t also strongly recommended to start the massage motors when the battery box used as
power.