User's Manual Part 2

102
Chapter 5 Capsule Endoscope Procedures
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
5.5 Initial setup of the recorder unit
The “Initial Setup of the Recorder Unit” function is not
available in Endo Capsule Software Light.
Connecting the recorder unit
The recorder unit is connected to the workstation via the recorder unit cradle.
If the antenna lead set being worn by a patient is connected
to the recorder unit, do not connect the recorder unit to the
workstation. Doing so may result in electric shock.
While setting up the recorder unit, do not remove the
recorder unit from the cradle or remove the cradle cable.
Doing so may result in loss of image data. If you do so,
connect the recorder to the workstation through the cradle
and the cradle cable and restart the workstation before
setting up the recorder unit again.
Up to two recorder units can be simultaneously connected to
the workstation.
If you connect a third recorder unit to the workstation, it will
be turned OFF.
Restart the workstation when the workstation does not work
properly.
Remove the USB devices from the workstation and restart
the workstation when it does not work properly. Then connect
the removed USB devices again.