User's Manual Part 2

106
Chapter 5 Capsule Endoscope Procedures
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Setting up the recorder unit
1.
Select the recorder unit for which the initial setup is to be performed.
If two recorder units are connected to the workstation, first
identify which recorder unit is being set up.
2. Click the [Initialize] button on the recorder management screen (see Figure
5.10). The initial settings screen is displayed.
Figure 5.10
When you perform the initial setup of the recorder unit, all
data on the recorder unit will be deleted. If you wish to keep
any of the data stored on the recorder unit, be sure to
download it before initializing.
If there are any image data remaining on the recorder unit, a
confirmation message will be displayed (see Figure 5.11). If
you click the [NO] button, the initial setup screen will be
displayed. Click the [YES] button to display the download
screen.
If the workstation cannot recogize the connected recorder(s),
please remove the cradle cable from the workstation, restart
the workstation and then connect it again.
Do not remove the recorder unit from the cradle while
initializing with two recorders connected whether the data is
now set to it or not. Doing so may result in the hangup of the
workstation. If the workstation does not work properly, please
restart the workstation.
Initialize button