User's Manual Part 2
Chapter 5 Capsule Endoscope Procedures
109
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
• Click the [Close detail] button to hide the details panel.
Figure 5.14
• The height, weight, and waist must be entered as a number
with 5 or less digits, with up to 3 digits in the integer part, and
up to 2 digits in the decimal part.
5. Click the [Initialize] button. The initial setup confirmation screen is displayed.
Figure 5.15
If you click the [Initialize] button with any essential
information missing, the message “Enter patient information
and examination date.” is displayed. Complete all the
essential information, and then click the [Initialize] button
again.
Close detail button
Initialize button