User's Manual Part 5

230
Chapter 7 Care Storage and Disposal
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Chapter 7 Care Storage and Disposal
7.1 Care
After wiping with a place of moistened gauze, dry the capsule
endoscope system thoroughly before using it again. If it is
used while still wet, there is the risk of electrical shock.
When cleaning the capsule endoscope system, always wear
appropriate personal protection equipment. Blood, mucous
and other potentially infectious material adhering the video
system center could pose and infection control risk.
Do not soak the system components in water, or sterilize
them using autoclave or gas. Doing so will damage the
system.
Do not wipe the external surfaces with hard or abrasive
wiping material. Doing so will scratch the surface.
After using the system, immediately perform the following cleaning procedure.
1. Remove dust, dirt, and other stains on the surface of the antenna pads and
the antenna lead set using a piece of gauze moistened with a neutral
detergent, then wipe the surfaces using a piece of gauze moistened with
70% ethyl or isopropyl alcohol.
2. Remove dust, dirt, and other stains on the surface of the following
equipment using a piece of gauze moistened with a neutral detergent, then
wipe the surfaces using a piece of gauze moistened with 70% ethyl or
isopropyl alcohol. After doing so, dry them thoroughly.
Do not soak the equipment in any liquid.
It may damage the equipment.
Recorder unit Recorder unit cradle Real time viewer
Real time viewer cable Recorder unit cradle cable Battery charger
Recorder unit harness Capsule activator