User's Manual Part 5

Chapter 8 Troubleshooting
235
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Recorder unit error messages
Error code on Recorder
Unit’s Display Panel
Possible Cause Solution
System error 1. Restart the system.
2. If the error persists, contact
Olympus.
RTC error 1. Connect the recorder unit to
the workstation, and perform
initialization.
2. If error persists after
initialization, contact Olympus.
Internal memory
error
Contact Olympus.
Battery pack error Replace with a new battery pack.
Internal memory full Connect the recorder unit to the
workstation, and download the
data from the Recorder Unit.
Uncharged battery
pack
Recharge the battery pack, or
replace it with a charged battery
pack.
Patient Information
not registered
Connect the recorder unit to the
workstation, and perform
initialization.
Low battery pack Recharge the battery pack, or
replace it with a charged battery
pack.
Less than 8 hours of
storage space left in
internal memory
Connect the recorder unit to the
workstation, and download the
data from the Recorder Unit.