Chapter 8 Troubleshooting
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
* The life span of a battery pack is approximately 1 year from the date of
purchase, or 320 recharging cycles, whichever is first. When the battery
pack approaches its end-of-life, the recorder unit will prompt for
replacement. Order a new battery pack and replace.
Real time viewer error messages
Workstation error messages
Time to replace
battery pack.
*
Order a new battery pack from
Oympus, and replace it. If you
continue to use the battery pack
for a prolonged period with these
warning icons are displayed, the
low battery pack icon ( on
page 235) may blink in
continuously, and may not allow
the examination to be performed.
Antenna lead set
disconnected
Connect the antenna lead set to
the recorder unit.
Error Code on Real Time
Viewer’s Display Panel
Possible Cause Solution
Uncharged battery
pack
Recharge the battery pack, or
replace it with a charged battery
pack.
E000 System error 1. Restart the system.
2. If the error persists, contact
Olympus.
Message Solutions
System error occurred. (E∗∗∗)
Contact service center.
Contact Olympus.
System is unstable.
Shutdown and restart workstation.
Shut down and restart workstation.
User ID and/or password is incorrect.
Enter correct user ID/password (case sensitive).
Enter correct user ID/password
(case sensitive) with the right
uppercase/lowercase combination.
Invalid value entered in date and/or time. Enter correct value.
Error code on Recorder
Unit’s Display Panel
Possible Cause Solution