User's Manual

OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL
105
Capsule endoscope
Product name Capsule endoscope
(EC-Y0005)
Field of view (maximum) 160°
Optics
Depth of field 0 – 20 mm
Sampling rate
2 fps
Power supply Power source Internal battery
Battery life 12 hours
Size Dimensions ø 11 (diameter) × 26 (length) mm
Type of protection
against electric shock
Internal power supply
Degree of protection
against electric shock
TYPE BF
Classification
as
medical
electrical
equipment
Degree of protection
against explosion
Use under combustible atmosphere
prohibited.
Protection
against fluid
ingress
IEC 60529 IPX-8
EMC Applied standard;
IEC 60601-1-2:2001
This instrument complies with the
standards listed in the left column.
CISPR 11 of emission:
Group 1, Class A
Year of
manufacturer
1712345
The year of manufacture is indicated
by the second digit of the serial
number.
The year of manufacture indicates the serial number of EC-Y0005.
FCC ID :S8QEC-S10 This device comlies with part 15 of
the FCC Rules. Operation is subject
to the following two conditions:
(1)This device may not cause
harmful interference, and (2) this
device must accept any interference
received, including interference that
may cause undesired operation.
IC :4763B-ECS10 This device complies with Industry Canada
licence-exempt RSS standard(s).
Operation is subject to the following
two conditions: (1) this device may
not cause interference, and (2) this
device must accept any
interference , including interference
that may cause undesired operation
of the device.