OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL
I
Inserting the Recorder Unit in the Recorder Unit Harness
• Attach the antenna unit and the recorder unit so that antenna cable is not caught on things
such as a knob of a door. Otherwise it may damage the antenna unit and/or the recorder
unit and may result in failure of the examination.
• Attach the antenna unit and the recorder unit so that antenna cable is not laid on or near
the antenna sheet. It may prevent the proper reception of images.
1. Insert the recorder unit into the pouch of recorder unit harness and close the pouch
cover. ( Refer to Figure 4.23) Please attach them so that an antenna cable is
between the body of the patient and the suspenders of the recorder unit.
Figure 4.23
4.3 Preparing the capsule endoscope
• Do not use capsule endoscopes that have been dropped, bitten, or been subjected to
excessive pressure. Doing so may result in an infection control risk as well as cause
injury to the patient due to damage to the capsule endoscope.
• Inspect the sterile package before you open it. If it is already opened or damaged, the
sterility of the capsule endoscope may have been compromised. Use a new capsule
endoscope instead.
• To prevent the capsule endoscope battery from deteriorating, store the capsule
endoscope under a temperature of 0 – 25°C (32 – 77°F). Do not store in a refrigerator or
a freezer, as condensation inside of the top cover may result.