User's Manual

OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL
91
Chapter 5 Care, Storage and Disposal
5.1 Care
After using the system, immediately perform the following cleaning procedure.
• After wiping with a piece of moistened gauze, dry this system
thoroughly before using it again. If it is used while still wet,
there is the risk of an electric shock.
• During cleaning, wear appropriate personal protective
equipment, such as eye wear, face mask, moisture-resistant
clothing and chemical-resistant gloves that fit properly and
are long enough so that the operator’s skin is not exposed.
Blood, mucous and other potentially infectious material
adhering to this system could pose an infection control risk.
• Do not soak the system components in water, or sterilize
them using autoclave or gas. Doing so will damage the
system.
• Do not wipe the external surfaces with hard or abrasive
wiping material. Doing so will scratch the surface.
• Do not wipe the connector of the recorder unit cradle with wet gauze.
A terminal becomes short and defective, and the charging the
recorder unit and downloading the data may not be possible.
1.
Remove dust, dirt, and other stains on the surface of the antenna unit
using a piece of gauze moistened with a neutral detergent, then wipe the
surfaces using a piece of gauze moistened with70% ethyl or isopropyl alcohol.
2. Remove dust, dirt, and other stains on the surface of the following
equipment using a piece of gauze moistened with a neutral detergent, then
wipe the surfaces using a piece of gauze moistened with 70% ethyl or
isopropyl alcohol. After doing so, dry them thoroughly.
• Recorder unit • Recorder unit cradle
• USB cable • Antenna harness
• Recorder unit harness • Capsule activator
Do not soak the equipment in any liquid.
It may damage the equipment.