Irregularity Possible cause Solution
The recorder unit is broken.
Contact Olympus.
The capsule endoscope is
broken.
Use a new capsule endoscope.
The antenna cable is
broken.
Contact Olympus.
The antenna unit is not
connected to the recorder
unit.
Connect the antenna unit to
the recorder unit properly.
Images from the
capsule
endoscope are not
displayed by the
Image View screen.
The antenna unit is too
far from the capsule
endoscope.
Reduce the distance between the
antenna and the
capsule endoscope.
The recorder unit
fails to turn OFF.
The power switch is not held
down long enough.
Hold down the power switch for
2 seconds or more.
Display layout is
out of place.
The PC’s DPI setting is not
set to 96DPI or 120DPI.
Right-click on the desktop and
select [Properties].
Select [Settings].
Click on the [Advanced] button.
Set [DPI setting] to either [96DPI]
or [120DPI].
You can not operate the
recorder unit by pushing any
button.
Troubles occur in software
of the recorder unit.
Remove the battery pack from the
recorder unit, and attach it again.
Then turn ON the recorder unit by
pushing the power button.
USB cable or AC adapter is
not connected normally.
Connect USB cable and AC
adapter once again.
WS does not recognize the
recorder unit
The USB cable or AC
adapter is broken.
Contact Olympus