User's Manual

30
4.1 Installing the ALT-Y0003 leak test air tubes and turning ON the power
ALT-Y0003 INSTRUCTION MANUAL
Ch.4
NOTE
To inform users of the time to replace the internal pump, “Caution Contact Olympus
for New Pump” will be lit up on the LCD monitor upon power activation. When this
indication is lit up, it is time to replace the pump. Turn OFF the power and contact
Olympus.
Disconnect the water-resistant cap from the ALT-Y0003 leak test air tube if
connected. Otherwise, a self-check error may occur.
1 Check that the outer surface of the ALT-Y0003 leak test air tubes is free from
scratches and flaws.
2 When connecting the equipment side connector of the ALT-Y0003 leak test air tube to
the tube connector, thoroughly wipe dust, dirt and stain out of the external surface of
the tube connector with a clean cloth.
3 Visually check that the inside of the removing grip of the ALT-Y0003 leak test air tubes
and the outside of the tube connector are free from water. If any water is observed,
wipe it with clean cloth.
4 Using one hand to hold the equipment in place,
hold the connecting grips of the ALT-Y0003
leak test air tube with the other hand and insert
them into the Tube connector 1 and Tube
connector 2 of this equipment until they stop.
Check that they do not come out by pulling
them slightly.
Figure 4.1
5 Set each ALT-Y0003 leak test air tube in the
tube hanger.
Figure 4.2
Tube connector 2
Tube connector 1
Connecting
grips