4.2 Self-check
ALT-Y0003 INSTRUCTION MANUAL
Ch.4
The self-check is performed to inspect for leakage within the channel and examine whether the pump
or valves works properly. This self-check is automatically performed when the power is turned ON, but
it can be also manually performed when necessary. For the manual self-check, refer to Section 7.3,
“Performing only self-check”.
PRECAUTION
Do not apply excessive force on the ALT-Y0003 leak test air tubes by forcibly
bending them or placing anything on it. Doing so may prevent a proper self-check
and damage the Leak Test Tubes.
Do not touch the ALT-Y0003 leak test air tubes during self-check. Otherwise, the
self-check cannot be conducted properly.
If an error code is displayed on the LCD monitor as a result of the self-check, follow
the instruction in Section 8.1, “Troubleshooting guide”.
NOTE
When self-check is complete, the results of self-check are automatically recorded in
this equipment.
4.2 Self-check
1 Turn ON this equipment. A self-check will
automatically start. The “Self Check In
Process” indicator will be displayed on the LCD
monitor. The number of mark “ ” of the
progress bar will increase to show the progress
of the self-check until all the process is
completed.
Figure 4.3
2 A short beep sounds when the self-check is completed properly. The “Scope ID
indicator (SCOPE1)” and “User ID indicator (USER)” indicators blinks as shown in
Figure 4.4, and the standby screen appears on the LCD monitor.
Figure 4.4