5.3 Preparation before use
ALT-Y0003 INSTRUCTION MANUAL
Ch.5
PRECAUTION
Use a basin deep enough to allow complete immersion of the endoscope and large
enough for the insertion section of the endoscope and the universal cord to be
coiled into a diameter of 17 cm or over. If the insertion section or universal cord is
coiled into a smaller diameter, the leakage testing may not be conducted properly.
Prepare equipment as shown below.
5.3 Preparation before use
Check Required items
This Automated Endoscope Leak Tester
ALT-Y0003 leak test air tubes (MAJ-2009) (× 2)
Personal protective equipment
A clean and large basin (Prepare two basins if two endoscopes are tested.)
Endoscope with a built-in Scope ID memory chip or a Scope ID tag
Water-resistant cap (only for EVIS series endoscope)
ID chip (MAJ-1546) (optional)
Printer (MAJ-1937) (optional)
ALT-Y0003 interface cable (MAJ-2052) (optional)