User's Manual

5.4 Power activation
39
ALT-Y0003 INSTRUCTION MANUAL
Ch.5
NOTE
When the printer is used, turn ON the printer before turning ON this equipment.
For indicators on the printer after power activation, refer to Table 4.1 in Section 4.1,
“Installing the ALT-Y0003 leak test air tubes and turning ON the power”.
5.4 Power activation
1 Confirm that ALT-Y0003 leak test air tubes are surely connected to Tube connector 1
and 2 of this equipment. Also confirm that an endoscope or no water resistance caps
are connected to the endoscope connectors of ALT-Y0003 leak test air tubes.
NOTE
An error code [E112] is displayed on turning power ON if the ALT-Y0003 leak test
air tubes are not securely connected to this equipment, or if an endoscope or water
resistance cap is connected to the ALT-Y0003 leak test air tubes.
2 Turn the power switch ON.
3 A self-check start automatically.
4 A short beep sounds when the self-check is completed properly. As seen in Figure
5.1, the Scope ID indicator (SCOPE1) and User ID indicator (USER) indicators blink.
The Standby screen is displayed on the LCD monitor.
Figure 5.1
5ECP5EQRG+&
ޓޓޓޓ1T7UGT+&
=/'07?)QVQ/GPW

SCOPE
1
SCOPE
2
USER