5.5 Recognition of scope ID
ALT-Y0003 INSTRUCTION MANUAL
Ch.5
NOTE
When the first endoscope is detected, the Endoscope ID indicator “SCOPE 1” lights
up.
When the second endoscope is detected, the Endoscope ID indicator “SCOPE 2”
lights up.
To replace the endoscope to be tested, press the STOP/BACK button and
disconnect the endoscope from the ALT-Y0003 leak test air tube. Go back to the
beginning of Section 5.5 and repeat the steps.
4 Check that the Tube connector 1 indicator is blinking. Connect an endoscope to the
ALT-Y0003 leak test air tube connected to Tube connector 1 according to the
instruction described in Section 5.6, “Connecting ALT-Y0003 leak test air tubes and
endoscope”.
NOTE
The error code [E122] is displayed when, in the step 4 above, an endoscope is
connected to the ALT-Y0003 leak test air tube that has been connected to Tube
connector 2. Press the Enter button and detach the endoscope from the
ALT-Y0003 leak test air tube. Go back to beginning of Section 5.5 and repeat the
steps.
Figure 5.5
5 To leak test a second scope, repeat steps 1 – 4 above, and connect the second
endoscope to the ALT-Y0003 leak test air tube connected to “Tube connector 2".