User's Manual
42
5.5 Recognition of scope ID
ALT-Y0003 INSTRUCTION MANUAL
Ch.5
NOTE
• When the first endoscope is detected, the Endoscope ID indicator “SCOPE 1” lights
up.
• When the second endoscope is detected, the Endoscope ID indicator “SCOPE 2”
lights up.
• To replace the endoscope to be tested, press the STOP/BACK button and
disconnect the endoscope from the ALT-Y0003 leak test air tube. Go back to the
beginning of Section 5.5 and repeat the steps.
4 Check that the Tube connector 1 indicator is blinking. Connect an endoscope to the
ALT-Y0003 leak test air tube connected to Tube connector 1 according to the
instruction described in Section 5.6, “Connecting ALT-Y0003 leak test air tubes and
endoscope”.
NOTE
The error code [E122] is displayed when, in the step 4 above, an endoscope is
connected to the ALT-Y0003 leak test air tube that has been connected to Tube
connector 2. Press the Enter button and detach the endoscope from the
ALT-Y0003 leak test air tube. Go back to beginning of Section 5.5 and repeat the
steps.
Figure 5.5
5 To leak test a second scope, repeat steps 1 – 4 above, and connect the second
endoscope to the ALT-Y0003 leak test air tube connected to “Tube connector 2".