5.7 Recognition of user ID
ALT-Y0003 INSTRUCTION MANUAL
Ch.5
To keep an endoscope leak test log, have this equipment recognize the individual User ID that
identifies the reprocessing operator.
NOTE
The START button on the main control panel does not function unless the User ID
has been recognized.
For addition or re-issue of the user ID card, contact Olympus.
If this equipment fails to detect the user ID, apply the provided user ID master card
to the RFID reader so that this equipment recognizes the user ID as the master ID.
If a problem with the user ID is observed, contact Olympus.
NOTE
When the User ID detection indicator does not illuminate on the control panel, the
user ID card has not been recognized. Bring the user ID card again close to the
RFID reader of equipment.
To replace the endoscope to be tested, press the STOP/BACK button and
disconnect the endoscope from the ALT-Y0003 leak test air tube. Go back to the
beginning of Section 5.5 and repeat the steps.
5.7 Recognition of user ID
1 Bring the user ID card of the leakage testing
worker close to the RFID reader of this
equipment until a short beep sounds.
Figure 5.8
2 Confirm that the User ID detection indicator on
the control panel is turned ON.
Figure 5.9