Use Instructions

20 Calibrate VS200 using the Olympus Calibration Slide
20.5 Shading Correction (Brightfield)
Use the stage navigator to move to the ‘empty’ area of the calibration slide.
The slide must be very clean (free of dust particles which will disrupt the
shading correction procedure).
1. In the [Calibrations] dialog box select the [Shading Correction] entry and
click the [Calibrate] button.
- 115 -
Instructions for useOLYMPUS VS200 System