Use Instructions

6. After the dark current image calibration select all non-immersion object-
ives. Immersion objectives shall be calibrated separately. Proceed with
[Next].
IMPORTANT
Actually, the label shading correction must only be performed for
the 2x objective. However you should do it for all other objectives
as well as otherwise there will always be an exclamation mark next
to the entry.
7. Use the stage navigator to move e.g. to position 2 (where you placed the
calibration slide for the label). Make sure you are in the center of the label
sticker.
8. Click the [Next] button to start the image acquisition for the 2x objective.
- 128 -
Instructions for use
20 Calibrate VS200 using the Olympus Calibration Slide
OLYMPUS VS200 System