Use Instructions

21 Additional calibrations for a fluorescence system
3. Use the stage navigator to move to a similar position on the calibration
slides to that shown in the image below.
4. Perform an autofocus.
5. If the focus differs too much you need to manually adjust the U-FFWO T3
adapter.
Adjust the camera adapter
1. Release the camera rotation locking screw. Hold the camera with one hand
to prevent camera rotation. Use the small wrench to adjust parfocality of
the camera until the image on the screen is in focus.
- 141 -
Instructions for useOLYMPUS VS200 System