Use Instructions

21 Additional calibrations for a fluorescence system
3. Select the 10x objective from the menu bar and perform an autofocus.
4. If the value is beyond the acceptable limits, loosen the screws in the beam
splitter flange to rotate the U-FFWO together with the camera.
5. After a slight rotation, tighten the screws again and click the [Back] button
in the dialog box.
6. Then click the [Autofocus] button in the next dialog box to perform the
test again.
7. Repeat until the value meets the standard.
8. Finalize the process by clicking the [Finish] button.