21.3 Camera shift
The camera shift calibration corrects the shift between the brightfield and
monochrome cameras. It is a software-based correction.
Measure the camera shift between the two cameras
1. Open the [Manual control] layout.
2. Display the cross hair by selecting the [Cross Hair] entry in the [View]
menu.
3. Select the 20x objective.