Use Instructions

21 Additional calibrations for a fluorescence system
2. Select all available objectives. Immersion are calibrated in the same wizard.
However the calibration wizard will calibrate all dry objectives first and sub-
sequently select the immersion medium objective(s).
3. Use the stage navigator to move the VS calibration slide to the correct
checkerboard area.
- 151 -
Instructions for useOLYMPUS VS200 System