Use Instructions

21 Additional calibrations for a fluorescence system
2. Click the [Calibrate] button.
The magnification test scan for the monochrome camera is identical
with the magnification test scan for the color camera. See Magnification
Test Scan auf Seite 136.
3. The resulting value should be within a deviation of +/- 0.1 degree.
In case the calibration value is not within the standard go back to the [Cam-
era-To-Stage Rotation] calibration and readjust the rotation. This will also
effect the shading correction, however. Refer to chapter Shading Cor-
rection (Brightfield) auf Seite 115 to redo the shading correction.
- 159 -
Instructions for useOLYMPUS VS200 System