5. Select all objectives. Immersion objectives are calibrated in the same wiz-
ard. However the calibration wizard will calibrate all dry objectives first and
subsequently select the immersion medium objective(s).
6. It is very important to select in the [Calibration method] group the cal-
ibration method [Special]. Only the calibration method [Special] will work
on a real sample. The calibration method [Standard] requires different
samples and is less effective. Also make sure that the checkbox [Shading
correction] is checked.
7. Proceed with [Next].
The system switches into live mode.
- 162 -
Instructions for use
21 Additional calibrations for a fluorescence system
OLYMPUS VS200 System