Use Instructions

21.8 Shading correction for darkfield (DFMono)
The darkfield observation method can only be used with 4x and 10x
objectives.
To perform shading correction for DFMono use the checkerboard sec-
tions on the VS-calibration slide.
1. In the [Calibrations] dialog box, select the [Shading Correction] entry and
click the [Calibrate] button.
2. Select the [DFMono] observation method and proceed with [Next].
3. Skip the acquisition of the dark current correction image. Proceed with
[Next].