20 Calibrate VS200 using the Olympus Calibration Slide
5. If the value is beyond the acceptable rotation angle, loosen the headless
hex screw used to fix the TV 0.63 adapter at the flange of the beam splitter.
6. After a slight rotation, tighten the screw again.
If the value is negative, tighten the screw counter clockwise.
If the value is positive, tighten the screw clockwise.
Click the [Back] button in the window. Then click the [Autofocus] button in
the next dialog box to perform the test again.
7. Adjust until the standard is met.
8. Finalize the process by clicking the [Finish] button.