Use Instructions

1. Remove the tray and switch to the [Manual control] layout.
2. Use the [Acquire] > [Devices] > [Device Settings] command to open the
[Device Settings] dialog box.
3. Select the [VS200L Multi-Sample Holder] > [Tray Types] entry in the tree
view.
4. Select the check box [Use Tray] next to the tray type you want to use.
24.4 Image too bright
If you see the following error message please adjust VS200 LED lamp voltages. See
ORCA camera adjustments auf Seite 91.
24.5 Setting the Koehler illumination
The units described below must be assembled and adjusted by Olympus. If these units are
assembled or adjusted by the customer, the operations are not ensured.
This chapter describes how to set the Koehler illumination.
Inbusschlüssel (Größe 2,5mm und 3mm mit Kugelkopf)
1. Open the door of the VS200 scanner.
2. Remove the lower tamper protection plate. To do so, loosen the 5 hex
- 180 -
Instructions for use
24 Troubleshooting