Use Instructions

24 Troubleshooting
Alternatively, you can also use the [Alt + F6] keystroke.
16. Release the field stop’s focus setting. To do so, loosen the hex screw (size
3mm) indicated in the figure.
17. Focus the field stop. To do so, insert the hex key into one of the holes indic-
ated in the figure. Move the hex key to the right and to the left to adjust the
focus ring.
Set the focus so that the edges of the closed field stop are sharp.
18. Tighten the field stops focus setting. To do so, use the hex screw (size 3
mm) indicated in the figure.
- 183 -
Instructions for useOLYMPUS VS200 System