7. The wizard will automatically select all of the objectives that are present and
calculate the correction factors.
If the Z parfocality for the 40x, 20x and 10x objectives is below
4µm everything is fine. The Z parfocality distance for the 2x object-
ive should not exceed 20µm.
If the values do not match the standard please check whether all
objective are screwed in completely and whether the calibration
slide is clean. Repeat the calibration.
8. Finalize the calibration by clicking the [Finish] button.
ATTENTION
In case you have calibrated an immersion medium objective as well clean the
objective and remove immersion medium residues from the calibration slide.
See Immersion Objectives auf Seite 38.