Use Instructions

Table Of Contents
11 Mounting fluorescence components
4. Connect the CAN-bus cable (which you find in the inside of the scanner) to
the lower connector and the CAN-terminator to the upper connector.
11.2.4 Add or replace filter (U-FFWO)
Hex key (3 mm)
1. Make sure that the power of the VS200 system is switched off.
2. To add or replace a 25 mm emission filter in the U-FFWO, open the cover of
the filter wheel using a 3 mm hex key.
3. Insert the filter into an empty position. Make sure that you orient the filter
correctly when you insert it.
- 51 -
Instructions for useOLYMPUS VS200 System