Use Instructions

Table Of Contents
Follow the safety precautions at all times during operation and maintenance of this
product. Non-observance may result in eye injury or damage to the system.
CAUTION
Risk of injury to the eyes
Do not look at operating lamp/LED as it can emit UV light.
Never look into the light emitting end of the light guide. The light could
severely damage the eye if the light is observed directly.
Always make sure the liquid light guide and X-Cite adapter are securely attached
to the VS200 system. This will minimize the risk of exposure to the UV light.
If the light source has a malfunction please contact Olympus or Excelitas cus-
tomer support. If the light source is serviced always make sure that the power
cord is disconnected.
Place the unit onto a hard, stable surface and make sure the ventilation openings
are not covered by something.
1. Unbox the LED light source and avoid bending or stretching of the liquid
light guide.
2. Mount the light guide into the X-Cite adapter and tighten the knurled
head screw.
Image
3. Insert the other end of the light guide into the LED light source (see man-
ufacturer’s operating manual for detailed information).
11.7 Filter set
The units described below must be assembled and adjusted by Olympus. If these units are
assembled or adjusted by the customer, the operations are not ensured.
Depending on the purchased filter set insert the single-band excitation filters into
the U-FFWR filter wheel. Refer to chapter Fluorescence filter wheels or camera
adapter for monochrome camera auf Seite 45 for the position information.
Insert the single-band emission filters into the U-FFWO (if available). Refer to
chapter Fluorescence filter wheels or camera adapter for monochrome camera
auf Seite 45 for position information.
Put the 25mm glass filter into position 8 of the U-FFWO.
Put the black-out filter into all empty filter wheel positions.
The following chapters describe how to mount filter cubes into the IX3-RFACA.
11.7.1 U-FF filter cube (IX3)
The units described below must be assembled and adjusted by Olympus. If these units are
assembled or adjusted by the customer, the operations are not ensured.
- 58 -
Instructions for use
11 Mounting fluorescence components
OLYMPUS VS200 System