Use Instructions

Table Of Contents
11 Mounting fluorescence components
1. Open the door of the VS200 scanner.
2. Slide the door of the IX3-RFACA to the left to insert an IX3 filter cube.
3. Insert a new filter cube into the correct position (position 2) as shown in the
image.
The correct position is printed inside of the IX3-RFACA and is hard
to read as the IX3-RFACA is mounted upside down. If you cannot
read it make sure that the filter is placed between the position
index 7 and 8 as shown in the image below.
4. After the cube is inserted slide the door of the IX3-RFACA back to the right.
11.7.2 Add or replace filter cube (IX3)
1. Make sure that the power of the VS200 system is switched off.
2. To add or replace an IX3 filter cube into the IX3-RFACA slide the door of
the IX3-RFACA to the left.
3. Insert the filter cube into an empty position.
4. Write down the position number of the filter cube as you will need it later.
5. Configure the mirror cube in the VS200 ASW software. To do this, open the
[Device Settings] dialog box.You can open this dialog box in the [Manual
control] layout.
6.
Click the [Previously Used Layout] button to go to a different layout.
You can find the [Previously Used Layout] button at the top right in the nav-
igation bar on the software's start page.
7. At the top right, on the menu bar click the [Manual control] button.
- 59 -
Instructions for useOLYMPUS VS200 System