18 VS200 device configuration
18.3 Device settings - objectives
A VS200 SLIDEVIEW kit is supplied with a 2x and a 20x objective ex-works. However,
the predefined device configuration already contains a 2x, 4x, 10x, 20x and 40x
objective.
You need to delete the objectives that are not installed from the nosepiece settings
in the [Device Settings] dialog box.
Open the [Device Settings] dialog box
1.
Use the [Previously Used Layout] button to go to a different layout. You
can find the [Previously Used Layout] button at the top right in the nav-
igation bar on the software's start page.
2. At the top right, on the menu bar click the [Manual control] button.
3. Select the [Acquire] >[Devices] > [Device Settings] command to open the
[Device Settings] dialog box.
Configuring objectives
1. Select the [IX3 nosepiece] entry on the left side and set all objectives that
are not installed to the status [Free].
Beispiel
Before
After
2. To add a phase contrast objective select the [Magnification] entry and the
correct [Objective Type] entry for e.g. position 5 and add e.g. [PH] in the
[Description] field to distinguish between ‘normal’20x and ‘PH 20x.
Beispiel