User Manual
Specifications
29
Page
5 Specifications
VS120-S1 VS120-S5 VS120-L100
Observable
specimens
Tissue specimen fixed on a slide glass with cover glass
Compatible slide
glass
Width: 25 to 26 mm, length: 75 to 76 mm, thickness: 0.8 to 1.4 mm
Specimens to be
observed
Compatible cover
glass
Thickness: 0.12 to 0.17 mm
Observation method Transmitted light brightfield observation (Koehler illumination)
Objective lens Automatic switching between 2X, 10X, 20X, and 40X objective
lenses
Motorized stage Automatic control via motorized XY stage
Specially-designed
optical microscope
Focusing Auto focus control via motorized focusing
Specially-designed
digital camera
Image correction Shading correction, white-balance function
Slide transfer
system
Number of slide glass
specimens
1 slide
(setting onto the
stage directly)
A maximum of 5
slides
(setting onto the
stage directly)
A maximum of 100
slides
(transferring onto the
stage by using the
slide loader)
Scan area
Width: 26 mm, length: 64 mm (slide glass 26 mm wide 76 mm
long)
Resolution
0.33 m/pixel or less when the 20X objective lens is used
Scan
Scan speed
About 2 minutes (area of 15 mm 15 mm when the 20X objective
lens is used)
controller DOS/V compatible controller (Windows 7 Professional English
version)
Network Ethernet (100/1,000)
Memory 4GB
Storage 1.0 TB or more
Display 24-inch wide-screen LCD monitor
Image saving format (proprietary format, JPEG, TIFF), a zoomed-in
image can be observed during image acquisition, annotation
function, automatic recognition of the position of a specimen, Z
stack function, EFI function, one-shot image acquisition function,
stepless scaling function, multiple image synchronized display
function, gluing together of images, consultation software (optional)
System control
Software
- Control of 5-piece
slide image
acquisition
Control of the
automatic transfer
system
Rated voltage/current Refer to page 31.
Weight About 52 kg
(including the weight
of the main body,
controller, and
monitor)
About 52 kg
(including the weight
of the main body,
controller, and
monitor)
About 100 kg
(including the weight
of the main body,
controller, and
monitor)
Operating
environment
Temperature: 15 to 28C, humidity: 30 to 80% (no condensation)
Other
Electric power
consumption
960W 960W 1030W