4.7.4  When in the menu mode, the display initially shows the current value of the 
parameters, such as the Setpoint of the control relay, while putting all of the outputs 
on hold. The two arrow buttons are used to adjust the display value up or down. 
To accept the new value press ENTER twice. While the value on the display is 
being changed, the relay outputs and the analog outputs remain on hold. 
The items that appear in the operation menu are:       
 •  Password  
 •  % Sat/ppm  
 •  Test  
 •  Calibration  
 •  Temperature  
 •  Status  
 •  Output High  
 •  Output Low 
 •  Relay A Setpoint 
 •  Relay A Deadband 
 •  Relay B Setpoint 
 •  Relay B Deadband 
 •  Alarm Relay High 
 •  Alarm Relay Low  
  4.8   DIP Switches  
4.8.1  The controls which are frequently used in the normal operation of the instrument are 
all accessible on the control panel. Some switches, which are infrequently used are 
located on the back of the swing-out board.         
4.8.2  The DIP switches are scanned on RESET, power-up and every time the instrument 
is taken into the menu mode. Therefore, after changes to the DIP switch settings, 
you must take the unit offline by pressing CALL in order for the instrument to scan 
the new DIP switch values.  
4.8.3  The following table describes the use of the 16 DIP switches:     
 DIP Switch
 Switch Switch 
 Bank S1
  Description of Use OFF ON 
  1  Selects temperature unit  °F  °C 
  2  Enables use of Password Feature  YES  NO 
  3  Fail Safe mode for Relay A  NO  YES 
  4  Direction of Control Relay A  Rising  Falling 
  5  Alarm Relay to Activate for Memory Loss  YES  NO 
  6  Fail Safe mode for Alarm Relay  NO  YES 
  7  Auto Return from menu if no button pressed  NO  YES 
  8  0-5 Vdc / 0-1 mA selector  Temperature  D.O. 