Oxygen Equipment User Manual

 Page 18 
  6.3   Reset    
6.3.1  The instrument can be reset without losing calibration or any of the stored values by 
pressing and releasing the S9 reset button (located at the bottom of the board near 
the center) This action is equivalent to turning the power off and on.     
  6.4   Output Hold  
6.4.1  It may be useful during some system maintenance procedures to place the relay 
and analog outputs on hold. To accomplish this simply press CALL. To return to 
on line operation press RUN. 
NOTE:  To safeguard against the operator forgetting to press RUN the instrument 
will automatically go back on line ten minutes after the last button was 
pressed provided this feature has been enabled by placing DIP switch 
No.7, Bank S1, in the ON position.         
7.0 UTILITY MENU  
  7.1   Utility Menu Functions  
7.1.1  The Utility Menu enables authorized personnel to perform the following: 
• Adjust the temperature output range.         
• Adjust the offset and span of the 0-1 mA/ 0-5 Vdc output 
• Adjust the offset and span of the 4-20 mA output   
• Perform the altitude compensation   
7.2   Access to Utility Menu  
7.2.1  The Utility Menu is protected by password. To access the Utility Menu press and 
hold both RUN and ENTER for five seconds until the PASSWORD LED illuminates. 
The RUN LED will flash to warn that Utility Menu is in use. Now with the arrow keys 
make the display read the password, "7". Press ENTER.        
7.2.2  The Utility Menu is entirely separate from the Operation Menu but uses the same 
LED display. Press CALL to step through the menu items.     
      Below is the cross-reference between menu items:       
  Display LED
  Meaning in Utility Menu 
  Output High  Temperature Output 100% (5 Vdc) 
  Output Low  Temperature Output 0% (0 Vdc) 
  %Sat/ppm  0-1 mA / 0-5 Vdc Output Adjust Low 
  Test  0-1 mA / 0-5 Vdc Output Adjust High 
  Calibration  4-20 mA Output Adjust Low 
  Temperature  4-20 mA Output Adjust High 
 Status  Altitude Compensation 