omega.com TM ® OMEGAnet® Online Service www.omega.com Internet e-mail info@omega.com Servicing North America: USA: ISO 9001 Certified Canada: One Omega Drive, P.O. Box 4047 Stamford CT 06907-0047 TEL: (203) 359-1660 e-mail: info@omega.com 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 e-mail: info@omega.
Omega FMG-3000 Series Magmeter 1. 2. 3. 4. 5. Topic: Quick Start Guide Specifications Installation Pipe Fittings Selecting a Location FMG-3000 Magmeter Configuration Page 1 2 3 3 4 6. 7. 8. 9. 10. 11. Topic: General Installation & Grounding Tips Wiring the Magmeter with 4-20 Loop out Wiring the Magmeter with freq. out Calibration Data Maintenance & Troubleshooting Ordering Information Page 5 5 6 6-10 11 12 1.
SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confirm chemical compatibility before use. 3. Do not exceed maximum temperature/pressure specifications. 4. Wear safety goggles or face shield during installation/service. 5. Do not alter product construction. 6. Disconnect power before attempting any service or wiring. 2. Specifications Power Requirements • 4 to 20 mA: • • 21.6 to 26.4 VDC, 22.1 mA maximum 400 mV p-p maximum ripple voltage Frequency: 4.5 to 26.
3. Installation: Pipe fittings Omega offers a wide selection of installation fittings that control the position of the Magmeter electrodes in relation to the dimensions of the pipe. You will find a complete list of order numbers for installation fittings in the Calibration tables on pages 9-13. Type Description Type Description Plastic tees 0.5 to 4 inch versions PVC or CPVC Iron, Carbon Steel, 316 SS Threaded tees 0.5 to 2 in.
5. FMG-3000 Magmeter Configuration Whether using the FMG-3000 (frequency or 4-20 mA output) the wiring terminals located on the inside of the yellow cover are identical. All of the connections from the Magmeter to external equipment (Datalogger, Chart Recorder, Flow meter, etc.) are made at the large 4-position terminal connector. When the cover is removed the wiring from the sensor can be seen connected to the smaller terminal block.
6. General Installation and Grounding Tips Sensor conditioning The Magmeter output signal may be unstable immediately after installation. Allowing the sensor to soak in a full pipe (or in any container of water) for 24 hours will stabilize the performance. • Very low conductivity fluids may require a longer conditioning period. (The Magmeter may not operate properly in fluids where the conductivity is less than 20 µS/cm.) Grounding The FMG-3000 Magmeter is unaffected by moderate levels of electrical noise.
. Wiring the FMG-3001 and FMG-3101 with Frequency output • • The FMG-3001 and FMG-3101 Magmeters output an open collector frequency signal. A 10 KΩ pull up resistor must be connected between terminals 1 and 2 if the magmeter is used with third-party equipment..
. Calibration Data: K-factors* and Full Scale Current Values Plastic Installation Fittings: PVC Tees and Saddles PIPE SIZE (IN.) FITTING TYPE K-Factor Gallons K-Factor Liters* 20 mA= in GPM 20 mA= in LPM SCH 80 PVC TEES FOR SCH 80 PVC PIPE ½ FP-5305 2289.37 ¾ FP-5307 1430.41 1 FP-5310 876.86 1¼ FP-5312 447.06 1½ FP-5315 324.19 2 FP-5315 206.69 2½ FP-5325 131.46 3 FP-5330 82.52 4 FP-5340 44.78 604.85 377.92 231.67 118.11 85.65 54.61 34.73 21.80 11.83 13.10 20.97 34.21 67.10 92.54 145.15 228.20 363.
9. Calibration Data: K-factors* and Full Scale Current Values Plastic Installation Fittings for Metric Pipes: Polypropylene True Union Tees and Wafers PVDF True Union Tees and Wafers PVC True Union Tees and Wafers PIPE SIZE (IN.) 8 FITTING TYPE K-Factor Gallons K-Factor Liters* 20 mA= in GPM 20 mA= in LPM POLYPROPYLENE FITTINGS (DIN/ISO ,BS, ANSI) ½ FP-5105PO 2192.73 ¾ FP-5107PO 1327.81 1 FP-5110PO 737.16 1¼ FP-5112PO 453.46 1½ FP-5115PO 275.03 2 FP-5120PO 156.87 579.32 350.81 194.76 119.81 72.
9. Calibration Data: K-factors* and Full Scale Current Values Metal Installation Fittings: Carbon Steel Tees and Weld-o-Lets Stainless Steel Tees and Weld-o-Lets Galvanized Iron Tees PIPE SIZE (IN.) FITTING TYPE K-Factor Gallons K-Factor Liters* 20 mA= in GPM 20 mA= in LPM 1572.66 1086.73 582.34 377.48 267.79 167.85 415.50 287.11 153.86 99.73 70.75 44.35 19.08 27.61 51.52 79.48 112.03 178.73 72.20 104.49 194.99 300.81 424.02 676.48 STAINLESS STEEL TEES ON SCH 40 PIPE ½ FMG-5305 1601.
9. Calibration Data: K-factors* and Full Scale Current Values Metal Installation Fittings: Bronze and Copper Tees and Brazolets PIPE SIZE (IN.) FITTING TYPE K-Factor Gallons K-Factor Liters* 20 mA= in GPM 20 mA= in LPM 582.34 330.54 254.76 157.36 153.86 87.33 67.31 41.58 51.52 90.76 117.76 190.64 194.99 343.53 445.71 721.58 COPPER TEES FITTING ON COPPER PIPE SCH K ½ FP-5305CU 2459.19 ¾ FP-5307CU 1108.02 1 FP-5310CU 649.87 1¼ FP-5312CU 422.03 1½ FP-5315CU 281.43 2 FP-5320CU 136.02 649.72 292.
10. Maintenance The FMG-3000 series Magmeter requires very little maintenance. There are no user-serviceable components in the Magmeter. • If the fluid contains deposits and solids that may coat the electrodes, a regular cleaning schedule is recommended. • Do not use abrasive materials on the metal electrodes. Clean with soft cloth and mild detergent only. • Use a cotton swab and mild detergent to remove deposits on the metal electrodes at the tip of the sensor. 10.1.
11.
MAD E IN USA SA WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation.